Test Pack and Bowie & Dick (BD125X1, BD125X2, BD125X3) KPCD-220, KPCD-222, PCD26-1

Documentation

Download BD125X10

Chemdye® Bowie & Dick test pack chemical indicator for Vacuum-assisted steam sterilization processes

Type 2. Applicable regulations: ISO 11140-1, ISO 11140-4.
Initial color: Yellow. Final color: Dark Brown-Black.
Exposure conditions: 10 min. 121°C. Size: 125 x 125 x 16 mm. Units per box: 20.

Bionova® Steam Test Pack PCD Kit. Process Challenge Device for Gravity air displacement steam or Vacuum-assisted steam sterilization processes

Spore: Geobacillus stearothermophilus (ATCC 7953). in a 106 population.
Applicable regulations: EN 285, ISO 11140-1, ISO 11140-4, ISO 11140-5.
Exposure conditions: 30 min. 121°C or 4 min. 132/134°C. Units per box: Kit containing 25 PCDs + 25 biological indicators.

KPCD-220: Bionova® Steam Test Pack PCD Kit. Process Challenge Device for Gravity air displacement steam or Vacuum-assisted steam sterilization processes

Spore: Geobacillus stearothermophilus (ATCC 7953). in a 106 population.
Applicable regulations: EN 285, ISO 11140-1, ISO 11140-4, ISO 11140-5.
Exposure conditions: 30 min. 121°C or 4 min. 132/135°C. Units per box: Kit containing 25 PCDs + 25 biological indicators.

PCD26-1: Integron® Emulator Indicator self-adhesive for Gravity air displacement steam or Vacuum-assisted steam sterilization processes

Spore: Geobacillus stearothermophilus (ATCC 7953). Type 5. Applicable regulations: EN 285, ISO 11140-1, ISO 11140-4.
Initial color: Yellow. Final color: according to reference.
Exposure conditions: 30 min. 121°C or 4 min. 132/134°C. Units per box: 25 PCDs or Individual PCD.